Proline Medical Splat Washing Process Cleaning Indicator - Very High Challenge Level

PROPLMCDWA4

Packet/200

Regular price
$297.00
Regular price
$297.00
Sale price
$297.00
Title

The Chemdye CDWA4 Splat Washing Indicator is designed for routine monitoring of cleaning processes and evaluating the performance of automatic and ultrasonic washing equipment.

Featuring a very high challenge level, the indicator helps detect the factors that may affect cleaning performance. The fuchsia-coloured organic test substance simulates difficult-to-remove contamination, allowing users to visually assess the effectiveness of a completed washing cycle.

The CDWA4 is suitable for monitoring automatic washing machines and ultrasonic washing machines operating at frequencies of 35 kHz or higher and temperatures between 30°C and 70°C.

Key Features

  • Designed for routine cleaning performance monitoring
  • Very high cleaning challenge level
  • Suitable for automatic and ultrasonic washing machines
  • Helps identify factors affecting the outcome of a cleaning cycle
  • Fuchsia organic test substance provides clear visual results
  • Synthetic, temperature-stable indicator support
  • Suitable for dental, healthcare and pharmaceutical applications
  • Can be evaluated visually using the supplied reference guide
  • Compatible with the optional Trazanto quality control and traceability system
  • Supplied in a pack of 200 indicators

Product Brochure

Instructions for Use

Product Specifications

Brand: Chemdye
Product Code: CDWA4
Product Type: Washing and cleaning performance indicator
Process: Cleaning and thermodisinfection
Challenge Level: Very high
Initial Colour: Fuchsia
Result: Removal of the fuchsia organic test substance
Indicator Size: Approximately 70 mm × 30 mm
Ultrasonic Frequency: 35 kHz or higher
Operating Temperature Range: 30°C to 70°C
Quantity: 200 indicators per pack
Standards: ISO 13485 and ISO/TS 15883-5
Applications: Dental, healthcare and pharmaceutical

Directions for Use

  1. Select the appropriate Chemdye Splat CDWA indicator for the required challenge level.
  2. Place the indicator into a compatible Chemdye Splat CDWAH or CDWAH-U holder.
  3. Place the assembled indicator and holder on the washing tray with the instruments or materials being processed.
  4. Remove the indicator once the cleaning cycle has finished.
  5. Compare the indicator with the reference guide supplied in the instructions for use and record the result.

If an unfavourable result is obtained, take the appropriate corrective action and repeat the washing process.

Always refer to the product instructions for use and the applicable facility procedures when interpreting results.